Electrical DC/Low Frequency
ให้บริการสอบเทียบเครื่องมือวัดทางด้านไฟฟ้า
Digital Multimeter DC Watt Meter
DC/AC Power Supply Withstanding Voltage Tester
Insulation Tester Process Calibrator
Oscilloscope Temperature Indicator/Simmulator/Transducers
AC/DC Volt/Ohm/Amp meter Inciruit Tester
Earth Resistance Tester Electrostatic Field Meter
LCR Meter Gauss Meter/ Tesla Meter
Analog MΩ HiTester Decade Resistance Box
Wrist Strap Tester Clamp Meter
Decade Resistance Box Power Meter
Magnet/Magnetic Flux Density Hight Volt Tester
Process Meter Profile Inspection Sensor
Profile Inspection Sensor Programmable DC Electronic Load
Puncture Tester RTD Calibrator
Static Meter Surface Resistance Meter
Etc. (Please contact)
|
Electrical – DC/Low Frequency ISO/IEC 17025 Accreditation |
|||
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
|
1 DC Voltage – Source |
(0 to 100) mV (> 0.1 to 1) V (> 1 to 10) V (> 10 to 100) V (> 100 to 1 000) V |
70 µV/V + 3.5 µV 58 µV/V + 5.8 µV 59 µV/V + 58 µV 64 µV/V + 0.58 mV 64 µV/V + 1.8 mV |
Multiproduct Calibrator |
|
1 DC Current - Source |
Up to 3.3 mA (3.3 to 33) mA (33 to 330) mA (0.33 to 2.2) A (2.2 to 11) A |
0.16 mA/A + 59 nA 0.12 mA/A + 0.3 µA 0.17 mA/A + 3.9 µA 0.35 mA/A + 52 µA 0.7 mA/A + 0.39 mA |
Multiproduct Calibrator |
|
1 DC Current – Source Clamp-On Ammeter |
(11 to 550) A |
5.8 mA/A + 0.18 A 5.9 mA/A + 1 A |
Multiproduct Calibrator with Coil |
|
1 AC Current – Source Clamp-On Ammeter |
@ 45 to 1 kHz (11 to 50) A (> 50 to 550) A |
6.6 mA/A + 0.3 A 7.6 mA/A + 1.2 A |
Multiproduct Calibrator with Coil |
|
1 AC Voltage – Source |
@ 10 Hz to 45 Hz (1 to < 33 mV (33 to < 330 mV (0.33 to < 3.3 V (3.3 to 33 V @ > 45 Hz to 10 kHz (1 to < 33 mV (33 to < 330 mV (0.33 to < 3.3 V (3.3 to < 33 V @ > 10 kHz to 20 kHz (1 to < 33 mV (33 to < 330 mV (0.33 to < 3.3 V (3.3 to < 33 V (33 to < 330 V @ > 20 kHz to 50 kHz (1 to < 33) mV (33 to < 330) mV (0.33 to < 3.3) V (3.3 to < 33) V @ > 50 kHz to 100 kHz (1 to < 33) mV (33 to < 330) mV (0.33 to < 3.3) V (3.3 to < 33) V @ > 100 kHz to 500 kHz (1 to < 33) mV (33 to < 330) mV (0.33 to < 3.3) V @ >45 Hz to 1 kHz (33 to < 330) V (330 to 1 000) V @ > 1 kHz to 10 kHz (33 to < 330) V @ > 1 kHz to 5 kHz (330 to 1 000) V @ > 5 kHz to 10 kHz (330 to 1 000) V |
4.1 mV/V + 24 µV 2.9 mV/V + 58 µV 1.8 mV/V + 0.29 mV 1.8 mV/V + 0.29 mV
2.4 mV/V + 24 µV 0.58 mV/V + 24 µV 0.35 mV/V + 70 µV 0.47 mV/V + 0.71 mV
2.9 mV/V + 24 µV 1.2 mV/V + 24 µV 0.93 mV/V + 70 µV 0.93 mV/V + 3.1 mV 1.1 mV/V + 39 mV
2.9 mV/V + 24 µV 1.9 mV/V + 47 µV 1.7 mV/V + 35 mV 2.2 mV/V + 5.8 mV
4.1 mV/V + 39 µV 2.8 mV/V + 0.2 mV 2.8 mV/V + 2 mV 2.8 mV/V + 20 mV
12 mV/V + 70 µV 8.1 mV/V + 0.39 mV 5.8 mV/V + 3.9 mV
0.58 mV/V + 7.7 mV 0.58 mV/V + 1 V
0.93 mV/V + 18 mV
2.4 mV/V + 0.12 V
2.4 mV/V + 0.28 V |
Multiproduct Calibrator |
|
1 AC Current - Source |
@ 10 Hz to 20 Hz (0.029 to ˂ 0.33) mA (0.33 to ˂ 3.3) mA (3.3 to ˂ 33) mA (33 to ˂ 330) mA @ > 20 Hz to 45 Hz (0.029 to ˂ 0.33) mA (0.33 to ˂ 3.3) mA (3.3 to ˂ 33) mA (33 to ˂ 330) mA @ > 45 Hz to 1 kHz (0.029 to ˂ 0.33) mA (0.33 to ˂ 3.3) mA (3.3 to ˂ 33) mA (33 to ˂ 330) mA (0.33 to ˂ 2.2) A @ > 1 kHz to 5 kHz (0.029 to ˂ 0.33) mA (0.33 to ˂ 3.3) mA (3.3 to ˂ 33) mA (33 to ˂ 330) mA (0.33 to ˂ 2.2) A @ > 5 kHz to 10 kHz (0.029 to ˂ 0.33) mA (0.33 to ˂ 3.3) mA (3.3 to ˂ 33) mA (33 to ˂ 330) mA @ 10 Hz to 45 Hz (0.33 to ˂ 2.2) A @ 45 Hz to 65 Hz (2.2 to 10) A @ > 65 Hz to 500 Hz (2.2 to 10) A @ > 500 Hz to 1 kHz (2.2 to 10) A |
2.9 mA/A + 0.18 µA 2.4 mA/A + 0.36 µA 2.4 mA /A + 3.6 µA 2.4 mA/A + 36 µA
1.5 mA/A + 0.18 µA 1.2 mA/A + 0.36 µA 1.2 mA /A + 3.6 µA 1.2 mA/A + 35 µA
1.5 mA/A + 0.28 µA 1.2 mA/A + 0.37 µA 1.1 mA /A + 3.6 µA 1.1 mA/A + 37 µA 1.2 mA/A + 0.39 mA
15 mA/A + 0.18 µA 2.4 mA/A + 0.36 µA 1.1 mA /A + 3.6 µA 1.1 mA/A + 37 µA 1.2 mA/A + 0.39 mA
15 mA/A + 0.18 µA 7 mA/A + 0.37 µA 7 mA /A + 3.6 µA 7 mA/A + 35 µA
2.4 mA/A + 0.38 mA
0.71 mA/A + 2.4 mA
1.2 mA/A + 2.4 µA
3.9 mA/A + 2.4 µA |
Multiproduct Calibrator |
|
1 Resistance - Source |
0 Ω (> 0 to ˂ 11) Ω (11 to 33) Ω (33 to 110) Ω (110 to ˂ 330) Ω (0.33 to ˂ 1.1) kΩ (1.1 to ˂ 3.3) kΩ (3.3 to ˂ 11) kΩ (11 to ˂ 33) kΩ (33 to ˂ 110) kΩ (110 to ˂ 330) kΩ (0.33 to ˂ 1.1) MΩ (1.1 to ˂ 3.3) MΩ (3.3 to ˂ 11) MΩ (11 to ˂ 33) MΩ (33 to ˂ 110) MΩ (110 to 330) MΩ |
9.3 mΩ 140 µΩ/Ω + 9.3 mΩ 142 µΩ/Ω + 17 mΩ 106 µΩ/Ω + 17 mΩ 105 µΩ/Ω + 17 mΩ 105 µΩ/Ω + 70 mΩ 105 µΩ/Ω + 70 mΩ 105 µΩ/Ω + 0.7 Ω 128 µΩ/Ω + 0.7 Ω 110 µΩ/Ω + 0.7 Ω 105 µΩ/Ω + 0.7 Ω 175 µΩ/Ω + 64 Ω 176 µΩ/Ω + 72 Ω 0.7 mΩ/Ω + 0.78 kΩ 1.16 mΩ/Ω + 1.7 kΩ 5.8 mΩ/Ω + 7 kΩ 5.9 µΩ/Ω + 0.2 MΩ |
Multiproduct Calibrator |
|
|||||||
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
||||
|
1DC Power – Source 33 mV to 1 020 V @ 3.3 mA to ˂ 9 mA @ 9 mA to ˂ 33 mA @ 33 mA to ˂ 90 mA @ 90 mA to ˂ 330 mA @ 0.33 A to ˂ 0.9 A @ 0.9 A to ˂ 2.2 A @ 2.2 A to ˂ 4.5 A @ 4.5 A to 11 A |
0.1089 mW to < 9 W 0.297 mW to ˂ 33 W 1.089 mW to ˂ 90 W 2.970 mW to ˂ 330 W 10.89 mW to ˂ 900 W 29.7 mW to ˂ 2 200 W 72.6 mW to ˂ 4 500 W (1.485 to 11 000) W |
0.47 mW/W + 58 mW 0.35 mW/W + 58 mW 0.47 mW/W + 76 mW 0.35 mW/W + 76 mW 0.35 mW/W + 76 mW 0.7 mW/W + 0.76 W 1.4 mW/W + 0.76 W 1.1 mW/W + 0.76 W |
Multiproduct Calibrator |
||||
|
1AC Power – Source (45 to 65) Hz 0.33 V to ˂ 1 020V @PF=1 @ 3.3 mA to ˂ 9 mA @ 9 mA to ˂ 33 mA @ 33 mA to ˂ 90 mA @ 90 mA to ˂ 330 mA @ 0.33 A to ˂ 0.9 A @ 0.9 A to ˂ 2.2 A @ 2.2 A to ˂ 4.5 A @ 4.5 A to 11 A |
1.089 mW to ˂ 9 W 2.97 mW to ˂ 33 W 10.89 mW to ˂ 90 W 29.7 mW to ˂ 330 W 108.9 mW to ˂ 900 W 297 mW to ˂ 2 200 W 726 mW to ˂ 4 500 W (1.485 to 11 000) W |
3 mW/W + 7.6 mW 1.8 mW/W + 7.6 mW 3 mW/W + 76 mW 1.8 mW/W + 76 mW 3 mW/W + 76 mW 1.8 mW/W + 0.76 W 2.4 mW/W + 0.76 W 1.8 mW/W + 0.76 W |
Multiproduct Calibrator |
||||
|
1 Capacitance – Source |
@50 Hz to 10 kHz (0.33 to ˂ 0.5) nF (0.5 to ˂ 1.1) nF (1.1 to ˂ 3.3) nF (3.3 to ˂ 11) nF (11 to ˂ 33) nF (33 to ˂ 110) nF (110 to ˂ 330) nF @50 Hz to 5 kHz 0.33 µF to ˂ 1.1 µF @50 Hz to 2 kHz (1.1 to ˂ 3.3) µF @50 Hz to 1.5 kHz (3.3 to 11) µF @ (50 to 800) Hz (11 to 33) µF @ (50 to 400) Hz (33 to 110) µF @ (50 to 200) Hz (110 to 330) µF @ (50 to 150) Hz (0.33 to 1) mF |
5.9 mF/F + 12 pF 5.9 mF/F + 12 pF 5.9 mF/F + 13 pF 5.9 mF/F + 13 pF 3 mF/F + 0.13 nF 3 mF/F + 0.13 nF 3 mF/F + 0.68 nF
3 mF/F + 1.3 nF
4.1 mF/F + 6.8 nF
4.1 mF/F + 14 nF
4.7 mF/F + 68 µF
5.9 mF/F + 0.14 µF
8.2 mF/F + 0.68 µF
12 mF/F + 0.68 µF |
Multiproduct Calibrator |
||||
|
1 Electrical Simulation of Thermocouple Indicators – Source |
Type B (600 to 800) °C (800 to 1 000) °C (1 000 to 1 550) °C (1 550 to 1 820) °C Type E (-250 to -100) °C (-100 to -25) °C (-25 to 350) °C (350 to 650) °C (650 to 1 000) °C Type J (-210 to -100) ºC (-100 to -30) ºC (-30 to 150) ºC (150 to 760) ºC (760 to 1 200) ºC |
0.32 °C 0.3 °C 0.27 °C 0.28 °C
0.41 °C 0.15 °C 0.13 °C 0.15 °C 0.18 °C
0.23 °C 0.15 °C 0.13 °C 0.15 °C 0.2 °C |
Multiproduct Calibrator |
||||
|
Electrical Simulation of Thermocouple Indicators - Source |
Type K (-200 to -100) °C (-100 to -25) °C (-25 to 120) °C (120 to 1 000) °C (1 000 to 1 372) °C Type N (-200 to -100) °C (-100 to -25) °C (-25 to 410) °C (410 to 1 300) °C Type R (0 to 250) °C (250 to 400) °C (400 to 1 000) °C (1 000 to 1 767) °C Type S (0 to 250) °C (250 to 1 000) °C (1 000 to 1 400) °C (1 400 to 1 767) °C Type T (-250 to -150) °C (-150 to 0) °C (0 to 120) °C (120 to 400) °C |
0.23 °C 0.15 °C 0.13 °C 0.15 °C 0.2 °C
0.32 °C 0.19 °C 0.17 °C 0.23 °C
0.48 °C 0.30 °C 0.28 °C 0.33 °C
0.41 °C 0.31 °C 0.38 °C 0.28 °C
0.52 °C 0.21 °C 0.15 °C 0.13 °C |
Multiproduct Calibrator |
||||
|
|||||||
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
||||
|
Electrical Simulation of RTD Indicators - Source |
Pt 385, 100 Ω (-200 to -80) °C (-80 to 0) °C (0 to 100) °C (100 to 300) °C (300 to 400) °C (400 to 630) °C (630 to 850) °C |
0.082 °C 0.082 °C 0.1 °C 0.12 °C 0.13 °C 0.16 °C 0.28 °C |
Multiproduct Calibrator |
||||
|
Electrical Simulation of RTD Indicators - Source |
Pt 3916, 100 Ω (-200 to -190) °C (-190 to -80) °C (-80 to 0) °C (0 to 100) °C (100 to 260) °C (260 to 300) °C (300 to 400) °C (400 to 600) °C (600 to 630) °C |
0.2 °C 0.32 °C 0.04 °C 0.047 °C 0.055 °C 0.063 °C 0.071 °C 0.078 °C 0.18 °C |
Multiproduct Calibrator |
||||
|
1 Electrical Simulation of Thermocouple Indicators - Measure |
Type B (600 to 800) °C (800 to 1 000) °C (1 000 to 1 550) °C (1 550 to 1 820) °C Type E (-250 to -100) °C (-100 to -25) °C (-25 to 350) °C (350 to 650) °C (650 to 1 000) °C Type J (-210 to -100) ºC (-100 to -30) ºC (-30 to 150) ºC (150 to 760) ºC (760 to 1 200) ºC Type K (-200 to -100) °C (-100 to -25) °C (-25 to 120) °C (120 to 1 000) °C (1 000 to 1 372) °C Type N (-200 to -100) °C (-100 to -25) °C (-25 to 120) °C (120 to 410) °C (410 to 1 300) °C |
0.38 °C 0.32 °C 0.27 °C 0.26 °C
0.30 °C 0.14 °C 0.13 °C 0.13 °C 0.13 °C
0.16 °C 0.15 °C 0.14 °C 0.14 °C 0.14 °C
0.23 °C 0.17 °C 0.16 °C 0.15 °C 0.16 °C
0.18 °C 0.17 °C 0.16 °C 0.15 °C 0.15 °C |
Digital Multimeter, 3458A |
||||
|
1 Electrical Simulation of Thermocouple Indicators - Measure |
Type R (0 to 250) °C (250 to 400) °C (400 to 1 000) °C (1 000 to 1 767) °C Type S (0 to 250) °C (250 to 400) °C (400 to 1 000) °C (1 000 to 1 767) °C Type T (-250 to -150) °C (-150 to 0) °C (0 to 120) °C (120 to 400) °C |
0.53 °C 0.32 °C 0.29 °C 0.25 °C
0.52 °C 0.33 °C 0.31 °C 0.29 °C
0.44 °C 0.18 °C 0.15 °C 0.14 °C |
Digital Multimeter, 3458A |
||||
|
Electrical Simulation of RTD Indicators - Measure |
Pt 385, 100 Ω (-200 to -80) °C (-80 to 0) °C (0 to 100) °C (100 to 300) °C (300 to 400) °C (400 to 630) °C (630 to 850) °C |
0.0068 °C 0.0068 °C 0.033 °C 0.035 °C 0.036 °C 0.039 °C 0.041 °C |
Digital Multimeter, 3458A |
||||
|
1 DC Voltage – Measure |
(0 to 100) mV (> 0.1 to 1) V (> 1 to 10) V (> 10 to 100) V (> 100 to 1 000) V |
18 µV/V + 0.36 µV 14 µV/V + 0.4 µV 14 µV/V + 0.7 µV 17 µV/V + 35 µV 28 µV/V + 0.12 mV |
Digital Multimeter, 3458A |
||||
|
1 AC Voltage - Measure |
@ 1 Hz to 40 Hz (1 to 10) mV (> 10 to 100) mV > 0.1 V to 1 V (> 1 to 10) V (> 10 to 100) V (> 100 to 1 000) V @ > 40 Hz to 1 kHz (1 to 10) mV (> 10 to 100) mV > 0.1 mV to 1 V (> 1 to 10) V (> 10 to 100) V (> 100 to 1 000) V |
0.35 mV/V + 3.5 µV 0.09 mV/V + 23.1 µV 0.09 mV/V + 47 µV 0.09 mV/V + 4.7 mV 0.24 mV/V + 4.7 mV 0.47 mV/V + 47 mV
0.24 mV/V + 1.3 µV 0.09 mV/V + 2.4 µV 0.09 mV/V + 24 µV 0.09 mV/V + 0.24 mV 0.24 mV/V + 2.4 mV 0.47 mV/V + 24 mV |
Digital Multimeter, 3458A |
||||
|
|||||||
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
||||
|
1 AC Voltage - Measure |
@ > 1 kHz to 20 kHz (1 to 10) mV (> 10 to 100) mV > 0.1 mV to 1 V (> 1 to 10) V (> 10 to 100) V (> 100 to 1 000) V @ > 20 kHz to 50 kHz (1 to 10) mV (> 10 to 100) mV > 0.1 mV to 1 V (> 1 to 10) V (> 10 to 100) V (> 100 to 1 000) V @ > 50 kHz to 100 kHz (1 to 10) mV (> 10 to 100) mV > 0.1 mV to 1 V (> 1 to 10) V (> 10 to 100) V (> 100 to 1 000) V @ > 100 kHz to 300 kHz (1 to 10) mV (> 10 to 100) mV > 0.1 mV to 1 V (> 1 to 10) V (> 10 to 100) V @ > 300 kHz to 1 MHz (10 to 100) mV > 0.1 mV to 1 V (> 1 to 10) V (> 10 to 100) V @ > 1 kHz to 2 MHz (10 to 100) mV > 0.1 mV to 1 V (> 1 to 10) V (> 10 to 100) V |
0.35 mV/V + 1.3 µV 0.17 mV/V + 2.4 µV 0.17 mV/V + 24 µV 0.17 mV/V + 0.24 mV 0.24 mV/V + 2.4 mV 0.7 mV/V + 24 mV
11.6 mV/V + 1.3 µV 0.35 mV/V + 2.4 µV 0.35 mV/V + 24 µV 0.35 mV/V + 0.24 mV 0.41 mV/V + 2.4 mV 1.4 mV/V + 24 mV
5.8 mV/V + 1.3 µV 0.93 mV/V + 2.4 µV 0.93 mV/V + 24 µV 0.93 mV/V + 0.24 mV 1.4 mV/V + 2.4 mV 3.5 mV/V + 24 mV
47 mV/V + 2.4 µV 0.35 mV/V + 12 µV 3.5 mV/V + 0.12 mV 3.5 mV/V + 1.2 mV 4.7 mV/V + 12 mV
12 mV/V + 12 µV 12 mV/V + 0.12 mV 12 mV/V + 1.2 mV 18 mV/V + 12 mV
18 mV/V + 12 µV 15 mV/V + 0.12 mV 18 mV/V + 1.2 mV 18 mV/V + 12 mV |
Digital Multimeter, 3458A |
||||
|
1 DC High Voltage – Measure |
(1 to 10) kV |
5.8 mV/V + 0.004 kV |
Kikusui 149-10A |
||||
|
1 DC High Voltage – Measure |
(1 to 20) kV |
5.8 mV/V + 0.11 V |
Tektronix TDS620A, Tektronix P6015 |
||||
|
1 AC High Voltage – Measure |
50 Hz (1 to 10) kV (1 to 20) kV |
12 mV/V + 0.006 kV 5.8 mV/V + 0.11 V |
Kikusui 149-10A |
||||
|
1 DC Current – Measure |
(0 to 100) µA (> 0.1 to 1) mA (> 1 to 10) mA (> 10 to 100) mA (> 0.1 to 1) A |
0.03 µA/A + 0.12 pA 0.04 nA/A + 8.2 nA 0.04 mA/A + 82 nA 0.06 mA/A + 0.9 µA 0.17 mA/A + 12 µA |
HP 3458A Multimeter |
||||
|
(> 1 to 3) A (> 3 to 10) A (> 10 to 20) A (> 20 to 30) A (> 30 to 50) A (> 50 to 100) A (> 100 to 200) A |
50 µA/A + 0.12 mA 55 µA/A + 1.2 mA 55 µA/A + 1.2 mA 3.5 mA/A 2.4 mA/A 2.4 mA/A 2.4 mA/A |
HP 3458A Multimeter, Standard Shunt |
|||||
|
1 AC Current – Measure |
@ 10 Hz to 20 Hz (10 to 100) µA (> 0.1 to 1) mA (> 1 to 10) mA (> 10 to 100) mA (> 0.1 to 1) A @ > 20 Hz to 45 Hz (10 to 100) µA (> 0.1 to 1) mA (> 1 to 10) mA (> 10 to 100) mA (> 0.1 to 1) A @ > 45 Hz to 100 Hz (10 to 100) µA (> 0.1 to 1) mA (> 1 to 10) mA (> 10 to 100) mA (> 0.1 to 1) A @ > 100 Hz to 1 kHz (10 to 100) µA @ > 100 Hz to 5 kHz (> 0.1 to 1) mA (> 1 to 10) mA (> 10 to 100) mA (> 0.1 to 1) A |
4.7 mA/A + 35 nA 4.7 mA/A + 0.25 µA 4.7 mA/A + 2.5 µA 4.7 mA/A + 25 µA 4.7 mA/A + 0.25 mA
1.8 mA/A + 35 nA 1.8 mA/A + 0.25 µA 1.8 mA/A + 2.5 µA 1.8 mA/A + 25 µA 1.9 mA/A + 0.25 mA
0.70 mA/A + 35 nA 0.70 mA/A + 0.25 µA 0.70 mA/A + 2.5 µA 0.70 mA/A + 25 µA 0.95 mA/A + 0.25 mA
0.70 mA/A + 35 nA
0.36 mA/A + 0.3 µA 0.36 mA/A + 2.4 µA 0.36 mA/A + 24 µA 1.2 mA/A + 0.24 mA |
HP 3458A Multimeter |
||||
|
1 AC Current – Measure |
@ > 5kHz to 20 kHz (> 0.1 to 1) mA (> 1 to 10) mA (> 10 to 100) mA (> 0.1 to 1) A @ > 20kHz to 50 kHz (> 0.1 to 1) mA (> 1 to 10) mA (> 10 to 100) mA (> 0.1 to 1) A @ > 50kHz to 100 kHz (> 0.1 to 1) mA (> 1 to 10) mA (> 10 to 100) mA |
0.7 mA/A + 0.3 µA 0.7 mA/A + 2.4 µA 0.7 mA/A + 24 µA 3.5 mA/A + 0.24 mA
4.7 mA/A + 0.5 µA 4.7 mA/A + 4.7 µA 4.7 mA/A + 47 µA 12 mA/A + 0.47 mA
6.4 mA/A + 1.8 µA 6.4 mA/A + 17.4 µA 6.4 mA/A + 0.18 mA |
HP 3458A Multimeter |
||||
|
|
|||||||
Parameter/Equipment |
Range |
Expanded Uncertainty of Measurement (+/-) |
Reference Standard, Method, and/or Equipment |
||||
|
1 AC Current – Measure |
(1 to 3) A, 50 Hz (3 to 10) A, 50/60 Hz (3 to 10) A, (0.3 to 1) kHz (3 to 10) A, (1 to 5) kHz (10 to 20) A, 50 Hz (20 to 30) A, 50 Hz |
0.012 A
0.041 A
0.039 A
0.043 A
0.077 A
0.112 A |
HP 3458A Multimeter, Agilent 34330A Shunt |
||||
|
1 Resistance – Measure |
(> 0 to 10) Ω (> 10 to 100) Ω (> 100 to 1000) Ω (> 1 to 10) kΩ (> 10 to 100) kΩ (> 0.1 to 1) MΩ (> 1 to 10) MΩ (> 10 to 100) MΩ (> 0.1 to 1) GΩ |
25 µΩ/Ω + 59 µΩ 22 µΩ/Ω + 0.59 mΩ 20 µΩ/Ω + 0.6 mΩ 20 µΩ/Ω + 7 mΩ 20 µΩ/Ω + 0.07 Ω 25 µΩ/Ω + 2.4 Ω 66 µΩ/Ω + 0.12 kΩ 0.59 mΩ/Ω + 1.2 kΩ 5.8 mΩ/Ω + 112 kΩ |
HP 3458A Multimeter |
||||
|
1 Insulation Resistance – Measure 100 V to 1 kV |
1 kΩ 10 kΩ 100 kΩ 1 MΩ 10 MΩ 100 MΩ 1 GΩ 10 GΩ 20 GΩ |
0.58 kΩ 0.6 kΩ 6 kΩ 58 kΩ 0.06 MΩ 0.59 MΩ 5.8 MΩ 59 MΩ 82 MΩ |
High Voltage Resistors Tinsley, ZX74, ZX84, TE-1M-20G |
||||
|
1 Resistance – Source |
(0.1 to 1) Ω (1 to 10) Ω (10 to 100) Ω (100 to 1 000) Ω (1 to 10) kΩ (10 to 100) kΩ 100 kΩ to 1 MΩ (1 to 10) MΩ (10 to 100) MΩ 1 GΩ 10 GΩ 20 GΩ |
24 mΩ 27 mΩ 13 mΩ 1.2 Ω 12 Ω 120 Ω 1.2 kΩ 13 kΩ 180 kΩ 58 MΩ 0.58 GΩ 1.2 GΩ |
Decade Resistor, Tinsley, ZX74, ZX84, Yokogawa, 2793-01, TE-1M-20G |
||||
|
1Inductance- Source 1kHz |
(1 to 10) µH (10 to 100) µH (0.1 to 1) mH (1 to 10) mH (10 to 100) mH (0.1 to 1) H (1 to 10) H |
0.59 µH 5.9 µH 59 µH 0.59 mH 5.9 mH 59 mH 0.59 H |
Inductor Box, BXL-70 |
||||
|
1Capacitance-Source @ 1 kHz |
(100 to 1000) pF (>1 to 10) nF (>10 to 100) nF (>100 to 1000) nF (>1 to 10) µF |
0.058 pF/pF 0.058 nF/nF 0.058 nF/nF 0.058 nF/nF 0.058 µF/µF |
Capacitor Box, BXC-05 |
||||
|
DC Magnetic Flux Density
1Tesla Meter, Gauss Meter
|
Up to 30 G (> 30 to 300) G (> 300 to 1 000) G (> 1 000 to 3 000) G (> 3 000 to 5 000) G (> 5 000 to 10 000) G |
0.6 G 0.7 G 1.3 G 5.7 G 7 G 12 G |
Standard magnet, Tesla Meter |
||||
|
DC Magnetic Flux Density
1Magnets |
Up to 300 G (> 300 to 500) G (> 500 to 3 000) G (> 3 000 to 5 000) G (> 5 000 to 10 000) G (> 10 000 to 15 000) G |
0.69 G 1.3 G 5.7 G 12 G 18 G 18 G |
Tesla Meter |
||||
1. On-site calibration service is available for this parameter, since on-site conditions are typically more variable than those in the laboratory, larger measurement uncertainties are expected on-site than what is reported on the accredited scope